Talbot- versus Shack-Hartmann Sensor: global aberration measurement in the presence of mid-frequency waviness

Johannes Schwider
  1. Institute of Optics, Information and Photonics, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU)

schwider [at] optik [dot] uni-erlangen [dot] de

Wave front sensors are applied in the measurement of the wave aberrations of the human eye. Such measurements have to guarantee a very high degree of safety being the basis for ablative corrections with the excimer laser. PMMA-platelets with an ablated surface profile are a possible material for a calibration retarder. Due to the point-like ablation process the resulting retarder shows a certain amount of mid-frequency waviness aggravating the measurement of global aberrations. It will be shown how this type of waviness hampers the performance of the Talbot and the Shack-Hartmann sensor. Fizeau-interferograms are used as a back-up.

Manuscript not yet submitted. The submission phase is currently closed.
@inproceedings{schwider2005a21,
  author     = {Schwider, Johannes},
  title      = {{Talbot- versus Shack-Hartmann Sensor: global aberration measurement in the presence of mid-frequency waviness}},
  booktitle  = {DGaO-Proceedings 2005},
  series     = {DGaO-Proceedings},
  eventtitle = {106. Jahrestagung der DGaO},
  eventdate  = {2005-05-17/2005-05-20},
  venue      = {Wrocław},
  publisher  = {Deutsche Gesellschaft für angewandte Optik e.V.},
  year       = {2005},
  pages      = {A21},
  langid     = {english},
  url        = {https://dgao-proceedings.de/paper/106-a21}
}
106. Annual Conference of the DGaO · Wrocław · 2005