Calibration of Si-SPAD detectors using the double attenuator technique and a low noise silicon photodiode

Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany
2 Cesky Metrologicky Institut (CMI), V Botanice 4, 15072 Praha 5, Czech Republic

marco.lopez@ptb.de

Abstract

Silicon single-photon avalanche diodes (Si-SPADs) are today the most frequently devices used for measuring ultra-weak optical flux in many quantum technology fields. Here the determination of the detection efficiency is a key parameter that has to be accurately known for achieving reliable measurements. At PTB the detection efficiency of Si-SPADs is determined by using the double attenuator technique, which uses a calibrated standard detector traceable to the primary standard. However, an alternative way to calibrate Si-SPADs is by direct comparison against a low optical flux transfer detector (LOFT-Detector). Such a calibration was recently carried out at PTB and compared to the one using the double attenuator technique. The calibration was carried out at a wavelength of 770 nm and at different optical power levels from 90 fW to 13 pW. The mean relative deviation of the detection efficiency determined by the LOFT-Detector and the double attenuator technique was < 1 %, thus within the combined standard uncertainty of the two measurements. Details of the setup and the calibration procedure will be discussed at the conference.

Keywords

Metrology Quantum Optics Radiometry
PDF herunterladen
@inproceedings{dgao117-p28, title = {Calibration of Si-SPAD detectors using the double attenuator technique and a low noise silicon photodiode}, author = {M. López, G. Porrovecchio, H. Hofer, B. Rodiek, M. Smid, S. Kück}, booktitle = {DGaO-Proceedings, 117. Jahrestagung}, year = {2016}, publisher = {Deutsche Gesellschaft für angewandte Optik e.V.}, issn = {1614-8436}, note = {Poster P28} }
117. Jahrestagung der DGaO · Hannover · 2016